GPRF Command Reference
Operating Manual 1202.3986.32 – 03 192
frequency/level list. The expected nominal power can be queried with the <List Index>
parameter; see example.
<List Index> Range: 0 to 499
Def. value: –
<Exp. Nom. Power> Range: –47 dBm to +34 dBm
Def. value: depending on list index (–10 dBm to –30 dBm in 10-
dB steps for index 0 to 19)
Example See Configuring the List Mode
Characteristics Firmware version V1.00
7.4.3.8 Measurement Results
The following commands return the RF power results of the measurement.
FETCh:GPRF:MEAS<i>:POWer[:RMS]:CURRent?
FETCh:GPRF:MEAS<i>:POWer:MINimum:CURRent?
FETCh:GPRF:MEAS<i>:POWer:MAXimum:CURRent?
FETCh:GPRF:MEAS<i>:POWer[:RMS]:AVERage?
FETCh:GPRF:MEAS<i>:POWer:PEAK:MINimum?
FETCh:GPRF:MEAS<i>:POWer:PEAK:MAXimum?
READ:GPRF:MEAS<i>:POWer[:RMS]:CURRent?
READ:GPRF:MEAS<i>:POWer:MINimum:CURRent?
READ:GPRF:MEAS<i>:POWer:MAXimum:CURRent?
READ:GPRF:MEAS<i>:POWer[:RMS]:AVERage?
READ:GPRF:MEAS<i>:POWer:PEAK:MINimum?
READ:GPRF:MEAS<i>:POWer:PEAK:MAXimum?
Returns RF power results, see Power: Results. The results of the current, average,
minimum and maximum traces can be retrieved. The additional mnemonics RMS,
MINimum, MAXimum denote the detector type.
Returned values
<Reliability>, Reliability indicator
<Power 1>,
...,
<Power n>
RF power for each measured power step. If the list mode is
switched off, a single value is returned (n = 1). In list mode, the
total number n of results is equal to the list count
(CONFigure:GPRF:MEAS<i>:POWer:LIST:COUNt?).
Range: –100.0 dBm to +57.0 dBm
Def. unit: dBm
Example See Single Shot and Continuous Power Measurements