CONFigure:GPRF:MEAS<i>:EPSensor:ATTenuation <Attenuation>
Specifies the external input attenuation factor to correct the power reading at the
sensor.
<Attenuation> Range: –50.00 dB to +50.00 dB
Def. value: 0.00 dB
Def. unit: dB
Example See Performing a Single Shot Measurement
Characteristics Firmware version V1.00
7.4.4.6 Measurement Results
The following commands return the results of the measurement.
FETCh:GPRF:MEAS<i>:EPSensor?
READ:GPRF:MEAS<i>:EPSensor?
Returns statistical evaluation values of the sensor power results, see External Power
Sensor: Results.
Returned values
<Reliability>, Reliability indicator
<Power Current>,
<Power Average>,
<Power Minimum>,
<Power Maximum>,
RF power (single values)
Range: depending on power sensor model
Def. unit: dBm
<Elapsed Statistics> Range: 1 to "statistic count" setting
(CONFigure:GPRF:MEAS<i>:EPSensor:SCOunt)
Example See Performing a Single Shot Measurement
Characteristics Firmware version V1.00
SCPI: Query only
For additional information concerning syntax elements and returned values refer to
External Power Sensor Measurement Commands.
7.4.5 Alphabetical List of Commands (GPRF)
Command Page
ABORt:GPRF:MEAS<i>:EPSensor...........................................................................................................195
ABORt:GPRF:MEAS<i>:POWer...............................................................................................................184
CALibration:GPRF:MEAS<i>:EPSensor:ZERO.........................................................................................197
CONFigure:GPRF:MEAS<i>:EPSensor:ATTenuation:STATe...................................................................198