CMW 500 GSM Applications
GSM Command Reference
Operating Manual 1202.3986.32 – 03 259
<Count> Number of measurement intervals for the "ACP Switching"
measurement
Range: 1 to 1000
Def. value: 10
Example See Specifying Measurement-Specific Settings
Characteristics Firmware version V1.00
CONFigure:GSM:MEAS<i>:MEValuation:SCONdition <Stop Condition>
Qualifies whether the measurement is stopped after a failed limit check or continued.
SONerror means that the measurement is stopped (STOP: ... MEAS<i>...) and
reaches the RDY state as soon as one of the results exceeds the limits.
<Stop Condition> NONE: Continue measurement irrespective of the limit check
SLFail: Stop measurement on limit failure
Def. value: NONE
Example See Specifying Measurement-Specific Settings
Characteristics Firmware version V1.00
CONFigure:GSM:MEAS<i>:MEValuation:MOEXception <Error Handling>
Defines whether the measurement shall be continued when an exception occurs.
<Meas on Exception> ON: continue
OF
F: stop
Def. value: OFF
Example See Specifying Measurement-Specific Settings
Characteristics Firmware version V1.00
CONFigure:GSM:MEAS<i>:MEValuation:MSLots
<Slot Offset>, <No of Slots>, <Measured Slot>
Defines settings for the measured slots.
<Slot Offset>, Start of the measurement interval relative to the GSM frame
boundary
Range: 0 to 7
Def. value: 0
<No of Slots>, Number of slots to be measured
Range: 1 to 8
Def. value: 8