CMW 500 WCDMA Applications
WCDMA GUI Reference
Operating Manual 1202.3986.32 – 03 319
On Limit Failure The measurement is stopped as soon as one of the limits is
exceeded, irrespective of the repetition mode set. If no limit
failure occurs, it is performed according to its "Repetition"
mode and "Statistic Length". Use this setting for
measurements that are essentially intended for checking
limits, e.g. production tests.
GPIB command:
CONFigure:WCDMa:MEASurement<i>:MEValuation:SCONdition
Measure on Exception
Specifies whether measurement results that the R&S CMW 500 identifies as faulty or
inaccurate are rejected. A faulty result occurs e.g. when an overload is detected. In
remote control, the cause of the error is indicated by the reliability indicator.
Off Faulty results are rejected. The measurement is continued;
the statistical counters are not re-set. Use this mode to
ensure that a single faulty result does not affect the entire
measurement.
On Results are never rejected. Use this mode e.g. for
development purposes, if you want to analyze the reason for
occasional wrong transmissions.
GPIB command:
CONFigure:WCDMa:MEASurement<i>:MEValuation:MOEXception
Multi Evaluation – Measurement Length
Defines the number of consecutive slots that form a single measurement interval; see
Defining the Scope of the Measurement. The measured slots are displayed in the
"Error Vector Magnitude", "Magnitude Error", and "Phase Error" diagrams.
GPIB command: CONFigure:WCDMa:MEASurement<i>:MEValuation:MSCount
Multi Evaluation – Preappointed Slot
Selects the slot to be used for the ACLR and spectrum emission measurement. See
"Defining the Scope of the Measurement" on page 303.
GPIB command: CONFigure:WCDMa:MEASurement<i>:MEValuation:PSLot
Multi Evaluation – Synchronization
Selects a slot number within the UL WCDMA frames (0 to 14) that the R&S CMW 500
will display as the first slot in the measurement interval. "Any" means that the
measurement is not synchronized to the UL WCDMA frame structure; the
measurement will start as fast as possible, beginning with the first captured slot.
GPIB command: CONFigure:WCDMa:MEASurement<i>:MEValuation:MSLot
Multi Evaluation – Modulation /CDP – Measurement Period
Selects a half-slot or a full-slot measurement.
Full Slot The modulation measurement results are based on the entire WCDMA