R&S
®
CMW 500 WCDMA Applications
WCDMA Command Reference
Operating Manual 1202.3986.32 – 03 334
9.4.2 WCDMA Multi Evaluation Measurement Commands
The commands valid for the WCDMA "Multi Evaluation" measurement are divided into
the groups listed below. For an overview of the commands refer to Command Groups.
Commands for control and configuration:
Measurement Control and States
Measurement Parameters
Statistical Settings (Modulation)
Statistical Settings (Spectrum)
Trigger Settings
Limits (Modulation)
Limits (Spectrum)
Commands for retrieval of measurement results:
EVM Measurement Results (Traces)
Magnitude Error Measurement Results (Traces)
Phase Error Measurement Results (Traces)
Spectrum Emission Measurement Results (Traces)
CDP vs Slot Measurement Results (Traces)
Spectrum Measurement Results
Measurement Results (Single Values)
CDP vs Slot Measurement Results (Single Values)
The following general measurement settings also affect the WCDMA multi-evaluation
measurement:
Signal Routing
Analyzer Settings
UE Signal Info
Issues of special interest for all commands:
MEAS<i> Abbreviation of "MEASurement<instance>". For better readability only
the abbreviated
form (which is also accepted by the instrument) is given
in the command reference.
It is possible to use up to four instances of the same firmware
application in parallel.
For further information refer to Firmware Applications.
Issues of special interest in the context of result retrieval:
FETCh and
READ
commands
Both commands can be used to retrieve measurement results:
FETCh... returns the results of the current measurement cycle
(single shot measurement) after they are valid. FETCh... must be
used after the measurement has been started (INITiate...,
measurement states RUN or RDY).
READ... starts a new single shot measurement and returns the
results.
For further information refer to Retrieving Measurement Results.