CMW 500 WCDMA Applications
WCDMA Command Reference
Operating Manual 1202.3986.32 – 03 341
<Resources State> QUE: measurement without resources, no results available
("queued")
ACT: resources allocated, acquisition of results in progress but
not complete ("active")
INV: not applicable because <main_state>: OFF or RDY
("invalid")
Example See Performing Single Shot Measurements
Characteristics Firmware version V1.00
SCPI: Query only
9.4.2.3 Measurement Parameters
The following commands define general settings for the multi-evaluation measurement.
CONFigure:WCDMa:MEAS<i>:MEValuation:REPetition <Mode>
Specifies the repetition mode of the measurement; see Statistical Settings. The
repetition mode specifies whether the measurement is stopped after a single shot or
repeated continuously. Use CONFigure:..:MEAS<i>:...:SCOunt to determine the
number of measurement intervals per single shot.
<Mode> SINGleshot: Single shot measurement
CONTinuous: Continuous measurement
Def. value: SINGleshot
Example See Single Shot and Continuous Measurements
Characteristics Firmware version V1.00
CONFigure:WCDMa:MEAS<i>:MEValuation:SCONdition <Stop Condition>
Qualifies whether the measurement is stopped after a failed limit check or continued.
SONerror means that the measurement is stopped (STOP: ... MEAS<i>...) and
reaches the RDY state as soon as one of the results exceeds the limits.
<Stop Condition> NONE: Continue measurement irrespective of the limit check
SLFail: Stop measurement on limit failure
Def. value: NONE
Example See Specifying Measurement-Specific Settings
Characteristics Firmware version V1.00