EasyManua.ls Logo

Rohde & Schwarz CMW 500 - Wimax Multi Evaluation Measurement Commands

Rohde & Schwarz CMW 500
450 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
R&S
®
CMW 500 WiMAX Applications
WiMAX Command Reference
Operating Manual 1202.3986.32 – 03 390
RF2C: RF 2 COM
Def. value: RF1C
Characteristics Firmware version V1.00
CONFigure:WIMax:MEAS<i>:RFSettings:EATTenuation <External Att>
Defines an external attenuation (or gain, if the value is negative), to be applied to the
selected RF input connector (ROUTe:...:MEAS<i>:RFSettings:CONNector).
<External Att> Range: –50.00 dB to +90.00 dB (increment 0.01 dB)
Def. value: 0.00 dB
Def. unit: dB
Characteristics Firmware version V1.00
10.3.2 WiMAX Multi Evaluation Measurement Commands
The commands valid for the WiMAX "Multi Evaluation" measurement are divided into
the groups listed below. For an overview of the commands refer to Command Groups.
Commands for control and configuration:
Measurement Control and States
Statistical Settings
Measurement Parameters
Zone Definition
Map of Bursts
Trigger Settings
Commands for retrieval of measurement results:
Error Vector Magnitude
I/Q Offset, Gain Imbalance, Quadrature Mismatch
Center Frequency Error, Burst Power, Crest Factor, CINR
Spectral Flatness
The following general measurement settings also affect the WiMAX multi-evaluation
measurement:
Analyzer Settings
Signal Routing
Issues of special interest for all commands:
MEAS<i> Abbreviation of "MEASurement<instance>". For better readability only
the abbreviated form (which is also accepted by the instrument) is given
in the command reference.
It is possible to use up to four instances of the same firmware
application in parallel.
深圳德标仪器 135-1095-0799

Table of Contents

Related product manuals