Remote Control
R&S
®
CMW500
174User Manual 1173.9463.02 ─ 02
identified in remote commands. For the status register of this application the mnemonic
UNIVersal is used (e.g. STATus:OPERation:TASK:A:GPRF:GENerator1:UNIVersal).
Signaling applications are not supported by the STATus:OPERation register.
Table 6-3: Bit definition for function group registers
Bit No. MEASurement GENerator
0 POWer, MEValuation UNIVersal
1 IQVSlot, PRACh not used
2 EPSensor, SRS not used
3 IQRecorder not used
4 FFTSanalyzer not used
The status registers at the lowest level indicate state transitions of the tasks as follows:
Table 6-4: Bit definition for application register
Bit No. MEASurement State GENerator State
0 OFF
This bit is set when the measurement is
switched off.
OFF
This bit is set when the generator is switched off.
1 QUED (queued)
This bit is set when the measurement is
queued.
PENDing
This bit is set when the generator has been switched
on but the output signal is not yet available.
2 RUN (running)
This bit is set when the measurement starts
running.
ON
This bit is set when the generator state changes from
"PENDing" to "ON".The generated signal is then
available at the output connector.
3 RDY (ready)
This bit is set when the (single-shot) mea-
surement has finished.
not used
4 SDReached (statistical depth reached)
This bit is set when the (continuous) mea-
surement has completed the first measure-
ment cycle (first shot).
not used
The reporting structures of the register hierarchy are administrable using the following
commands:
STATus:OPERation:TASK:A:ENABle etc.
STATus:OPERation:TASK:A:ESRQ etc.
STATus:OPERation:TASK:A:NTRansition etc.
STATus:OPERation:TASK:A:PTRansition etc.
The states and state transitions of a specific task can be queried using the following
commands:
STATus:OPERation:TASK:A:GPRF:GENerator1:UNIVersal:CONDition?
etc.
Status Reporting System