Basic Instrument Functions
R&S
®
CMW500
91User Manual 1173.9463.02 ─ 02
Last Calibration
Display the type and date of the last calibration of the instrument. Remote commands
allow also to query information about previous calibrations.
Possible types of calibration are:
●
F/S Correction: Correction performed in factory or service
●
User Correction: Correction performed by the user
●
Calibration: Verification in the factory
●
Outgoing Calibration: Verification by the service
SCPI command:
CALibration:BASE:ALL
CALibration:BASE:LATest
CALibration:BASE:LATest:SPECific
5.2.4.2 Problem Reporting
The "Problem Reporting" section of the "Info" dialog allows to prepare collected logging
information for problem reporting to Rohde & Schwarz.
In the case that you encounter problems with your instrument, proceed as follows to send
the generated log files to Rohde & Schwarz:
1. Press the utility key "Info" to open the info dialog.
2. Press the softkey "Problem Reporting". A list of logging sessions is displayed as
shown in the figure below.
3. If required (e.g. because of limitations in the electronic mail system) use the hotkey
"Max. File Size" to adjust the maximum size of the compressed log files to be trans-
ferred to Rohde & Schwarz.
4. Select the directory of the session where the problem occurred. The date and time
of the log session refer to the start of the instrument.
5. Press the hotkey "Prepare for sending". The selected log files are packed into one
or more compressed files located in folder "Output".
6. Send the files to Rohde & Schwarz for analysis. You can copy the files from folder
"Output" using the hotkey "Copy" or you can navigate to the directory "D:\Rohde-
Schwarz\CMW\Log\output" and copy the files stored there. To transfer the files from
the instrument to a PC you can e.g. transfer them via LAN or use a USB stick.
Utility Dialogs