MEAS Key - Receiver R&S ESCI
1166.6004.12 4.18 E-2
Multiple detection is important in EMI measurements since, for example, civil standards specify limits for
both the quasi-peak and the average value. Thanks to the multiple use of detectors, only one test run is
needed. The peak detector can be combined with any other detector since it is the fastest detector and
therefore ideal for overview measurements.
DETECTOR menu:
DETECTOR
CISPR
AVERAGE
MIN PEAK
MAX PEAK
QUASIPEAK
AVERAGE
QP RBW
UNCOUPLED
RMS
The DETECTOR softkey opens a submenu to select
the detector and combinations of detectors.
A maximum of three detectors can be switched on
simultaneously.
Note: This softkey is also available in the Receiver-
menu.
MAX PEAK
The MAX PEAK softkey activates the max peak detector.
IEC/IEEE-bus command DET:REC POS
MIN PEAK
The MIN PEAK softkey activates the min peak detector.
IEC/IEEE-bus command DET:REC NEG
QUASIPEAK
The QUASIPEAK softkey selects the quasi-peak detector.
The IF bandwidth is adapted as a function of the frequency range.
The coupling of the IF bandwidth to the frequency range can be
cancelled using softkey QP RBW UNCOUPLED.
IEC/IEEE-bus command DET:REC QPE
AVERAGE
The AVERAGE softkey activates the average detector.
IEC/IEEE-bus command DET:REC AVER