MEAS Key - Receiver R&S ESCI
1166.6004.12 4.28 E-2
Structure of the ASCII file:
RECEIVER mode, final measurement data:
Content of file Description
File header
Type; R&S
ESCI;
Version;2.32;
Date;03.Aug 2004;
Mode;Receiver;
Start;10000;Hz
Stop;100000;Hz
x-Axis;LIN;
Scan Count;1;
Transducer;TRD1;
Scan 1:
Start;150000;Hz
Stop;1000000;Hz
Step;4000;Hz
RBW;100000;Hz
Meas Time;0.01;s
Auto Ranging;ON;
RF Att;20;dB
Auto Preamp;OFF;
Preamp;0;dB
Instrument model
Firmware version
Date record storage date
Instrument operating mode
Start/stop of the display range.
Unit: Hz
Scaling of x axis linear (LIN) or logarithmic (LOG)
Number of scans set
Transducer name (if switched on)
Loop over all defined scan ranges (1-10)
Range - start frequency in Hz
Range – stop frequency in Hz
Range - step width in Hz for linear step width or in
% (1-100) for logarithmic step width
Range - resolution bandwidth
Range - measurement time
Auto ranging on - or off for current range
Range - input attenuation
Auto Preamp on or off for current range
Range - preamplifier on (20dB) or off (0dB)
Data section
of the file
TRACE 1 FINAL:
Trace Mode;CLR/WRITE;
Final Detector, QUASI PEAK;
x-Unit;Hz;
y-Unit;dBuV;
Final Meas Time;1.000000;s
Margin;6.000000:s
Values;8;
2;154000.000000;81.638535;
15.638535;N;GND
1;158000.000000;86.563789;
7.563789;N;GND
2;1018000.000000;58.68987
3;-1.310127;N;GND
...
Selected trace
Trace mode:
CLR/WRITE,AVERAGE,MAX HOLD,MIN HOLD,
VIEW, BLANK
Final detector
MAX PEAK, MIN PEAK, RMS, AVERAGE,
QUASI PEAK, AC VIDEO
Unit of x values:
Unit of y values:
Final measurement time
Margin
Number of test points
Measured values:
<Trace>;<x-value>, <y-value>; <phase>;
<ground>
Phase and protective grounding are output only if
a line impedance stabilization has been activated.
They specify the setting at which the maximum
RFI level at the associated frequency was found.