R&S ESCI Receiver – MEAS Key
1166.6004.12 4.35 E-2
THRESHOLD
ON OFF
The THRESHOLD ON OFF softkey activates or deactivates the
THRESHOLD SCAN measurement function. This function will also be
activated on opening the submenu with the THRESHOLD SCAN softkey
from the RECEIVER main menu.
EDIT PATH
DECIM SEP
. ,
NEW
APPEND
HEADER
ON OFF
ASCII
COMMENT
EDIT PEAK
LIST
EDIT PEAK
LIST
EDIT
FREQUENCY
INSERT
DELETE
SORT BY
FREQUENCY
SORT BY
DELTA LIM
ASCII
CONFIG
ASCII
EXPORT
ASCII
CONFIG
PAGE UP
PAGE DOWN
The EDIT PEAK LIST softkey calls the
EDIT PEAK LIST submenu used for
editing the peak list.
Further functions relating to the peak
list are described in the operating
manual, section "Data Reduction by
Generating Subrange Maxima".
NO OF
PEAKS
With the NO OF PEAKS softkey, you can enter the number of final
measurement peaks to be stored. Numbers between 1 and 500 can be
entered. If the selected number is attained, no further final measurements
will be performed.
IEC/IEEE-bus command :CALC:PEAK:SUBR 1...500
NB/BB
DISCR
With the NB/BB DISCR softkey, you can enter the decision threshold to
be used by the analyzer to discriminate between broadband and
narrowband interference. Values between 0 dB and 200 dB can be
entered.
IEC/IEEE-bus command :FMEasurement:NBBB ON | OFF
:FMEasurement:NBBB:LEVel 0..200dB
MARGIN
The MARGIN softkey activates the entry field of the margin, i.e. of an
additional acceptance threshold for the determination of the peak list. The
limit line currently used is shifted by this amount for defining the maxima.
The range of values is -200 dB to 200 dB.
IEC/IEEE-bus command :CALC:PEAK:MARG –200dB...200dB