MEAS Key - Receiver R&S ESCI
1166.6004.12 4.36 E-2
FINAL
MEAS TIME
The FINAL MEAS TIME softkey activates the entry field of the time of final
measurement.
IEC/IEEE-bus command :SWE:TIME:FME <num_value>
AUTOMATIC
FINAL
The AUTOMATIC FINAL softkey activates the automatic mode for the final
measurement, i.e. a final measurement will be performed automatically and
immediately each time a value out of limit is detected.
IEC/IEEE-bus command :FMEasurement:AUTO ON
INTER
ACTIVE
The INTERACTIVE softkey selects the following sequence for the final
measurement:
• The prescan is interrupted – HOLD SCAN state.
• The bar graph measurement is started in the free running mode.
• The signal can be exactly analyzed by modifying the receiver settings.
• The actual final measurement is started, the receiver settings being
restored except the frequency.
• The current frequency replaces the original one in the frequency list
(drifting interference sources).
• The prescan is continued at the frequency at which it was interrupted....
Note: With the AUTOMATIC FINAL softkey in the CONTINUE
FINAL MEAS submenu a switchover can be made to the
automatic mode before the measurement is started anew.
IEC/IEEE-bus command :FMEasurement:AUTO OFF
Sequence for AUTOMATIC FINAL:
RUN
SCAN
HOLD
SCAN
STOP
SCAN
STOP
FINAL MEAS
HOLD
FINAL MEAS
The RUN SCAN softkey starts the prescan. The HOLD SCAN submenu
is called.
If an out-of-limit value is detected, the receiver automatically goes to the
HOLD SCAN state and starts the final measurement. The HOLD FINAL
MEAS submenu comes up. On completion of the final measurement, the
receiver continues the prescan, and the HOLD SCAN submenu is
displayed again.