R&S ESCI Receiver – MEAS Key
1166.6004.12 4.37 E-2
Sequence for INTERACTIVE:
RUN
SCAN
HOLD
SCAN
STOP
SCAN
STOP
FINAL MEAS
SKIP
FREQUENCY
AUTOMATIC
FINAL
INTER
ACTIVE
GET
MAXHOLD
MEASURE
The RUN SCAN softkey starts the prescan. The HOLD SCAN submenu
is called.
If an out-of-limit value is detected, the receiver automatically goes to the
HOLD SCAN state. A submenu with several options for the final
measurement comes up:
• AUTOMATIC FINAL activates the automatic final measurement
mode for the rest of the test run.
• SKIP FREQUENCY skips the final measurement and continues with
the prescan.
• GET MAXHOLD accepts the highest level measured during the
HOLD SCAN state as the result of the final measurement and
continues the prescan. (The level value in question is displayed as a
small bar in the bar graph.)
• MEASURE starts the final measurement, the receiver settings being
restored except the frequency.
• STOP FINAL MEAS aborts the final measurement and the prescan.