R&S ESCI Receiver – MEAS Key
1166.6004.12 4.41 E-2
Receiver-User-Interface
PIN
N 14
L1 15
L2 16
L3 17
GND 12
PE fl. 18
+5 V 13
46 N
21 L1
22 L2
23 L3
50 GND
47 PE fl.
48 +5 V
ESH2-Z5
PIN
Filter
Fig. 4-3 Connection R&S ESCI with R&S ESH2-Z5 (for direct connection without a filter: cable EZ-
13)
Receiver-User-Interface
N 14
L 15
PE fl. 18
GND 12
+5 V 13
2 N
6 L
3 PE fl.
8 DGND
1 +5 V
ESH3-Z5
PIN
Filter
Pin
Fig. 4-4 Connection R&S ESCI with R&S ESH2-Z5 (for direct connection without a filter: cable EZ-
14, model 02)
Receiver-User-Interface
N 14
L1 15
L2 16
L3 17
GND 12
+5 V 13
14 N
15 L1
16 L2
17 L3
12 GND
13 +5 V
ENV 4200
PIN
Filter
Pin
Fig. 4-5 Connection R&S ESCI with R&S ENV4200 (for direct connection without a filter: cable EZ-
21)
For controlling the phase selection and the PE simulating network of the V-networks ESH2-Z5, ESH3-Z5
and ENV42000, the +5-V supply voltage and some control lines are to be routed through the wall of the
shielded room.
The connection cables EZ-14 and EZ-5 can be provided for the 4-line V-network ESH2-Z5, the
connection cable EZ-14 and EZ-6 for the two-line V-network ESH3-Z5 and the connection cables ES-14
and EZ-21 for the 4-line V-network ENV4200.