R&S ESCI Receiver - MKR->
1166.6004.12 4.59 E-2
SETTINGS
COUPLED
The SETTINGS COUPLED softkey couples the receiver frequency settings from
the corresponding subscans to the marker frequency for functions TUNE TO
MARKER and MARKER TRACK.
IEC/IEEE bus command :CALC:MARK:SCO ON
SEARCH
LIMITS
RIGHT
LIMIT
LEFT
LIMIT
THRESHOLD
SEARCH
LIMITS OFF
The SEARCH LIMITS softkey limits the search range for
maximum or minimum search. The softkey switches to a
submenu in which the search range limits can be set in the x
and y direction.
LEFT
LIMIT
RIGHT
LIMIT
The LEFT LIMIT and RIGHT LIMIT softkeys define the two
vertical lines SL1 and SL2. The search is performed between
these lines in the frequency and time domain
If only one line is enabled, line SL1 is the lower limit and the
upper limit corresponds to the stop frequency. If SL2 is also
enabled, it determines the upper limit.
IEC/IEEE-bus command: CALC:MARK:X:SLIM:LEFT 1MHZ
CALC:MARK:X:SLIM:RIGH 10MHZ
CALC:MARK:X:SLIM ON
THRESHOLD
The THRESHOLD softkey defines the threshold line.
The threshold line represents a limit for the level range of the
max. search at the lower end and that of the min. search at the
upper end.
IEC/IEEE-bus command: CALC:THR -20dBm
CALC:THR ON