R&S ESCI Receiver - TRACE
1166.6004.12 4.65 E-2
FINAL
MAX PEAK
The FINAL MAX PEAK selects the max peak detector for
the final measurement.
IEC\IEEE bus command :DET:FME POS
FINAL
MIN PEAK
The FINAL MIN PEAK selects the min peak detector for the
final measurement.
IEC\IEEE bus command :DET:FME NEG
FINAL
QUASIPEAK
The FINAL QUASIPEAK selects the quasi peak detector
for the final measurement.
IEC\IEEE bus command :DET:FME QPE
FINAL
AVERAGE
The FINAL AVERAGE selects the average detector for the
final measurement.
IEC\IEEE bus command :DET:FME AVER
FINAL
CISPR AV
The FINAL CISPR AV selects the weighting average
detector according to CISPR 16-1 for the final
measurement.
IEC\IEEE bus command :DET:FME CAV
FINAL
RMS
The FINAL RMS selects the rms detector for the final
measurement.
IEC\IEEE bus command :DET:FME RMS
PEAK LIST
ON OFF
The PEAK LIST ON / OFF softkey switches on and off the indication of the
peak list or of the final measurement results. The single value is indicated as
+ or x. The assignment of symbol to trace is fixed.
RUN SCAN automatically switches PEAK LIST to OFF in order to prevent the
indication of preceding final measurement results. PEAK SEARCH
automatically sets PEAK LIST to ON (see Data Reduction and Peak List)
IEC\IEEE bus command DISP:TRAC:SYMB CROS | OFF