R&S ESCI Receiver - TRACE
1166.6004.12 4.67 E-2
Example:
File contents Description
File header
Type;R&S
ESCI
Version;1.00;
Date;10.Nov 03;
Mode;Receiver;
Start;150000.000000;Hz
Stop;1000000000.000000;Hz
x-Axis;LOG;
Detector;AVERAGE;
Scan Count;1;
Transducer;;
Instrument model
Firmware version
Date of data set storage
Instrument mode
Start/stop of the display range.
Unit: Hz
Scaling of x-axis linear (LIN) or logarithmic (LOG)
Selected detector
MAXPEAK, MINPEAK, AVERAGE, RMS, QUASIPEAK
Scan count
Transducer name (if switched on)
Data section of the file
Scan ranges
Scan 1:
Start;150000.000000;Hz
Stop;30000000.000000;Hz
Step; 4000.000000;Hz
RBW;9000.000000;Hz
Meas Time;0.001000;s
Auto Ranging;OFF;
RF Att;10.000000;dB
Auto Preamp;OFF;
Preamp;0.000000;dB
Scan 2:
Start;30000000.000000;Hz
Stop;1000000000.000000;Hz
Step; 50000.000000;Hz
RBW;120000.000000;Hz
Meas Time;0.000100;s
Auto Ranging;OFF;
RF Att;10.000000;dB
Auto Preamp;OFF;
Preamp;0.000000;dB
Settings for scan range 1
Range 1 - start frequency in Hz
Range 1 – stop frequency in Hz
Range 1 - step width
Range 1 - resolution bandwidth
Range 1 - measurement time
Range 1 - Auto ranging on - or off
Range 1 - input attenuation
Range 1 - Auto Preamp on or off
Range 1 - preamplifier on (20dB) or off (0dB)
Settings for scan range 2
Range 2 - start frequency in Hz
Range 2 – stop frequency in Hz
Range 2 - step width
Range 2 - resolution bandwidth
Range 2 - measurement time
Range 2 - Auto ranging on - or off
Range 2 - input attenuation
Range 2 - Auto Preamp on or off
Range 2 - preamplifier on (20dB) or off (0dB)
Data section of the file
trace
Trace 1:
Trace Mode;CLR/WRITE;
x-Unit;Hz;
y-Unit;dBµV;
Values;26863;
150000.000000;15.604355;
154000.000000;13.236252;
158000.000000;11.907021;
...;...;
Selected trace
Trace mode: CLR/WRITE, MAXHOLD
Unit of x values: Hz for span > 0
Unit of y values:
dB*/V/A/W depending on the selected unit
Number of test points
Measured values:
<x value>, <y value>;