Level Display / RF Input R&S ESCI
1166.6004.12 4.76 E-1
Level Display Setting and RF Input Configuration – AMPT Key
The AMPT key is used to set the reference level, the maximum level and the display range of the active
window as well as the input impedance and the input attenuation of the RF input.
The AMPT key opens a menu for setting the reference level and the input attenuation of the active
window. The data entry for the reference level (REF LEVEL softkey) is opened automatically.
Further settings regarding level display and attenuation can be made in this menu.
REF
LEVEL
GRID
ABS REL
UNIT
RF ATTEN
AUTO
RANGE
LOG 100 dB
RANGE
LINEAR
RANGE
LOG MANUAL
REF LEVEL
POSITION
REF LEVEL
OFFSET
RF ATTEN
MANUAL
RF INPUT
50Ω 75Ω
EL ATTEN
AUTO
EL ATTEN
MANUAL
EL ATTEN
OFF
Option
FSP-B25
Option
FSP-B25
Option
FSP-B25
AMPT
dBm
dBµV
dBµA
VOLT
AMPERE
WATT
dBpW
dBmV
RANGE
LINEAR %
RANGE
LINEAR dB
REF LEVEL
The REF LEVEL softkey allows the reference level to be input in the currently
active unit (dBm, dBµV, etc).
IEC/IEEE-bus command: DISP:WIND:TRAC:Y:RLEV -60dBm
RANGE
LOG 100 dB
The RANGE LOG 100 dB softkey sets the level display range to 100 dB.
IEC/IEEE-bus command: DISP:WIND:TRAC:Y:SPAC LOG
DISP:WIND:TRAC:Y 100DB