Analyzer - Sweep R&S ESCI
1166.6004.12 4.90 E-1
Sweep Settings – SWEEP Key
The SWEEP key serves for configuring the sweep mode.
SWEEP menu
SWEEP
MEAS
BW
TRIG
CONTINUOUS
SWEEP
SWEEPTIME
AUTO
SWEEP
COUNT
SINGLE
SWEEP
SGL SWEEP
DISP OFF
SWEEPTIME
MANUAL
CONTINUE
SGL SWEEP
SWEEP
POINTS
The SWEEP key calls a menu in which the sweep
mode is defined. In split-screen mode, the entries
made are valid for the active window only.
The CONTINUOUS SWEEP, SINGLE SWEEP
and SGL SWEEP DISP OFF softkeys are mutually
exclusive selection keys.
CONTINUOUS
SWEEP
The CONTINUOUS SWEEP softkey activates the continuous sweep mode,
which means that the sweep takes place continuously according to the trigger
mode set.
When working in the split-screen mode and with different settings in the two
windows, screen A is swept first, followed by screen B. When the softkey is
pressed, the sweep is restarted.
CONTINUOUS SWEEP is the default setting of R&S ESCI.
IEC/IEEE-bus command: INIT:CONT ON
SINGLE
SWEEP
The SINGLE SWEEP softkey starts n sweeps after triggering. The number of
sweeps is determined by the SWEEP COUNT softkey.
When working in the split-screen mode, the frequency ranges of the two
windows are swept one after the other.
If a trace is swept using TRACE AVERAGE or MAXHOLD, the value set via
the SWEEP COUNT softkey determines the number of sweeps. If 0 has been
entered, one sweep is performed.
IEC/IEEE-bus command: INIT:CONT OFF