MKR -> - Analyzer R&S ESCI
1166.6004.12 4.144 E-1
PEAK: signal 1
NEXT PEAK RIGHT: signal 2
NEXT PEAK RIGHT: signal 4
NEXT PEAK RIGHT: no further signal detected
With 6 dB peak excursion, all signals will be detected with NEXT PEAK. With
NEXT PEAK RIGHT, all signals to the right of signal 1 will be detected.
Order of signals detected:
Order of signals detected:
PEAK: signal 1
NEXT PEAK: signal 2
NEXT PEAK: signal 3
NEXT PEAK: signal 4
or
PEAK: signal 1
NEXT PEAK RIGHT: signal 2
NEXT PEAK RIGHT: signal 4
NEXT PEAK RIGHT: no further signal detected
EXCLUDE
LO
The EXCLUDE LO softkey limits the frequency range for the marker search
functions or disables the limit.
activated Because of the feedthrough of the first local oscillator to the first
intermediate frequency at the input mixer, the LO is represented
as a signal at 0 Hz. To avoid the marker jumping to the LO at 0
Hz with the peak function when setting the display range, this
frequency is excluded. The minimum frequency to which the
marker jumps, is 6 × resolution bandwidth (RBW).
deactivated No restriction to the search range. The frequency 0 Hz is in-
cluded in the marker search functions.
IEC/IEEE-bus command: CALC:MARK:LOEX ON