R&S ESCI Analyzer - MEAS
1166.6004.12 4.147 E-1
POWER
ON
OFF
The POWER ON/OFF softkey switches the power measurement on and off.
When entering the submenu it is ON since the power measurement is already
switched on with the TIME DOM POWER softkey in the main menu.
Note: The measurement is performed on the trace on which marker 1 is
placed. To evaluate another trace, marker 1 should be set on
another trace using the SELECT TRACE softkey in menu MKR.
IEC/IEEE-bus command: CALC:MARK:FUNC:SUMM:PPE ON
CALC:MARK:FUNC:SUMM:PPE:RES?
CALC:MARK:FUNC:SUMM:RMS ON
CALC:MARK:FUNC:SUMM:RMS:RES?
CALC:MARK:FUNC:SUMM:MEAN ON
CALC:MARK:FUNC:SUMM:MEAN:RES?
CALC:MARK:FUNC:SUMM:SDEV ON
CALC:MARK:FUNC:SUMM:SDEV:RES?
PEAK
The PEAK softkey switches on the calculation of the peak value from the points of
the displayed trace or a segment thereof.
For the maximum peak, the largest peak value obtained since the activation of
MAX HOLD ON is displayed.
With AVERAGE ON, the peak values of a trace are averaged over several
sweeps and displayed.
The number of sweeps over which the average or the maximum value is
calculated is set with the NUMBER OF SWEEPS softkey.
IEC/IEEE-bus command: CALC:MARK:FUNC:SUMM:PPE ON
CALC:MARK:FUNC:SUMM:PPE:RES?
RMS
The RMS softkey switches on the calculation of the rms value from the points of
the displayed trace or a segment of it.
For the maximum peak, the largest rms value obtained since the activation of
MAX HOLD ON is displayed.
With AVERAGE ON, the rms values of a trace are averaged over several sweeps
and displayed.
The number of sweeps over which the average or the maximum value is
calculated is set with the NUMBER OF SWEEPS softkey.
IEC/IEEE-bus command: CALC:MARK:FUNC:SUMM:RMS ON
CALC:MARK:FUNC:SUMM:RMS:RES?