MEAS - Analyzer R&S ESCI
1166.6004.12 4.164 E-1
Examples:
1. Measurement of adjacent-channel power for a specific standard:
The adjacent-channel power is to be measured for a signal at 800 MHz with 0 dBm level in line with
IS136.
[
PRESET] Set the R&S ESCI to the default setting.
[
FREQ: CENTER: 800 MHz] Set the center frequency to 800 MHz.
[
AMPT: 0 dBm] Set the reference level to 0 dBm.
[
MEAS] Call the menu for the measurement functions.
[
CHAN PWR / ACP] Select the channel and adjacent-channel power measurement function.
The measurement is performed with the default settings or a previously
defined setting. The submenu for setting the desired new configuration
is opened.
[
CP/ACP STANDARD:
select IS136:
ENTER] Select the NADC (IS136) standard.
[
CP/ACP CONFIG] Call the submenu for configuration of the adjacent-channel power
measurement.
[
NO. OF ADJ CHAN:
2
ENTER] Select two adjacent channels for the measurement, i.e. the adjacent
channel and the alternate channel are measured.
[
ADJUST SETTINGS] Set the optimum span, resolution bandwidth (RBW), video bandwidth
(VBW) and detector automatically for the measurement. The absolute
channel power and the relative power of the adjacent channels are
displayed on the screen.
PREV
Change to the main menu for channel power measurement.
[
ADJUST REF LVL] Set the reference level equal to the channel power measured.