Limit Lines R&S ESCI
1166.6004.12 4.188 E-1
LINES - EDIT LIMIT LINE menu
EDIT LIMIT
LINE
NEW LIMIT
LINE
The EDIT LIMIT LINE and NEW LIMIT LINE softkeys both call the EDIT
LIMIT LINE sub-menu used for editing limit lines. In the table heading,
the characteristics of the limit line can be entered. The data points for
frequency/time and level values are entered in the columns.
Name Enter name.
Domain Select domain.
Unit Select units.
X-Axis Select interpolation
Limit Select upper and lower limit value.
X-Scaling Entry of absolute or relative values for the X axis
Y-Scaling Entry of absolute or relative values for the Y axis
Margin Entry of margin.
Threshold Entry of vertical threshold (only with relative vertical
scaling)
Comment Enter comments.
Time/Frequency Enter time/frequency for the data points.
Limit/dBm Enter magnitudes for the data points.
Note: Domain, unit, X scaling and Y scaling cannot be modified
as soon as reference values have been entered in the
data section of the table.
NAME
The NAME softkey enables the entry of characteristics in the table heading.
Name - Enter name
A maximum of 8 characters is permitted for each name. All names must be
compatible with the MS DOS conventions for file names. The instrument
stores all limit lines with the .LIM extension.
IEC/IEEE-bus command: CALC:LIM3:NAME "GSM1"
Domain - Select time or frequency domain
The default setting is frequency.
Note: A change in domain (frequency/time) is only permitted when the data
point table is empty.
IEC/IEEE-bus command: CALC:LIM3:CONT:DOM FREQ