R&S ESCI Configuration/Setup
1166.6004.12 4.213 E-3
Name/Unit/Interpolation: Cable dB LI
N
EDIT TRANSDUCER FACTOR
Comment:
FREQUENCY FREQUENCYTDF/dB.. TDF/dB..
1.0000000 MHz
1.0000000 GHz
1.000
5.500
The table with the data of the marked factor (the EDIT softkey) or an empty table is displayed in which
the following entries are preset (the NEW softkey):
Unit: dB
Interpolation: LIN for linear frequency scaling
LOG for logarithmic frequency scaling
The characteristics of the factor can be entered in the header field of the table, the frequency and the
transducer factor in the columns of the table.
Name Entry of the name
Unit Selection of unit
Interpolation Selection of interpolation
Comment Entry of a comment
FREQUENCY Entry of the frequency of the reference points
TDF/dB Entry of the transducer factor.
An overwritten transducer factor remains stored in the background as long as the edited factor is stored
with the SAVE TRD FACTOR softkey or until the table is closed. A factor overwritten by mistake can be
restored by leaving the entry.
Name - Entry of name
A maximum of 8 characters is permissible for the name. The characters
have to comply with the convention of DOS file names. The unit
automatically stores all transducer factors with the extension .TDF.
If an existing name is changed, the factor stored under the previous name is
retained and will not be overwritten automatically with the new name. The
previous factor can be deleted at a later time, using DELETE. Thus, factors
can be copied.
IEC/IEEE-bus command CORR:TRAN:SEL <name>