Configuration/Setup R&S ESCI
1166.6004.12 4.216 E-3
New Entry and Editing of Transducer Sets
A transducer set is characterized by:
− maximum 10 ranges (Ranges) in which different transducer factors can be active
− the combination of several transducer factors per range (Factor)
− a transducer-set name (Name)
NEW
EDIT
INS BEFORE
RANGE
RANGE
1-5 6-10
INS AFTER
RANGE
DELETE
RANGE
SAVE TRD
SET
The NEW and EDIT softkeys both open the submenu
for editing and entering new transducer factors if softkey
TRANSDUCER SET.
Name/Unit/Break dBuV OFF
EDIT TRANSDUCER SET
RANGE1 RANGE2
Start
Stop
Factors
Comment:
RANGE3
RANGE4 RANGE5
30 MHz
300 MHz
Cable
Antenna1
300 MHz
1 GHz
Cable
Preamp
Antenna2
The table with the data of the marked set (the EDIT softkey) or an empty table in which the following
entries are preset (the NEW softkey) is displayed:
Unit: dB
Break: OFF
The characteristics of the set can be entered in the header field of the table, the subranges in the
columns of the set.
Name Entry of the name
Unit Selection of unit
Break Activating the query when changing the subrange
Comment Entry of a comment
Start Entry of the start frequency of the subrange
Stop Entry of the stop frequency of the subrange
Factors Selection of the transducer factors for the subrange