R&S ESCI Configuration/Setup
1166.6004.12 4.217 E-3
An overwritten transducer set remains stored in the background as long as the edited factor is stored
with the SAVE TRD SET softkey or until the table is closed. A set overwritten by mistake can be
restored by leaving the entry.
Name - Entry of name
A maximum of 8 characters is permissible for the name. The characters have
to comply with the convention of DOS file names. The unit automatically
stores all transducer sets with the extension .TDS.
If an existing name is changed, the set stored under the previous name is
retained and will not be overwritten automatically with the new name. The
previous set can be deleted at a later time using DELETE FACTOR/SET.
Thus, sets can be copied.
IEC/IEEE-bus command CORR:TSET:SEL <name>
Unit - Selection of unit
The unit of the transducer set is selected from a selection box activated by
the ENTER key.
The unit should be selected prior to the entry as it determines the settable
transducer factors. The preset unit for new sets is "dB".
SET U
dBpT
IEC/IEEE-bus command CORR:TSET:UNIT <string>
Break - Activation of query when changing a subrange
The sweep can be stopped if the user changes the subrange and selects a
new subrange of the transducer. A message informs the user that the limit
has been attained. He can continue the sweep or switch off the transducer.
The interruption is activated by setting Break to ON. Selection is by the
ENTER key which toggles between ON and OFF (toggle function).
IEC/IEEE-bus command CORR:TSET:BRE ON|OFF
Start - Entry of start frequency of subrange
Stop - Entry of stop frequency of subrange
The individual subranges have to be linked without a gap. That is why the
start frequency is already defined from the second subrange (= stop
frequency of previous range).
IEC/IEEE-bus command CORR:TSET:RANG<1.10>
<freq>,<freq>,<name>..