R&S ESCI Configuration/Setup
1166.6004.12 4.223 E-3
Serial Interface Configuration
SETUP-GENERAL SETUP submenu:
COM
INTERFACE
The COM INTERFACE softkey activates the COM INTERFACE table for
entry of the serial interface parameters.
The following parameters can be configured in the table:
Baud rate data transmission rate
Bits number of data bits
Parity bit parity check
Stop bits number of stop bits
HW-Handshake hardware handshake protocol
SW-Handshake software handshake protocol
Owner assignment to the measuring instrument or computer
Baud
Bits
Parity
Stopbits
HW-Handshake
SW-Handshake
Owner
COM INTERFACE
8
None
1
None
XON/XOFF
Instrument
9600
Baud – Data transmission rate
The R&S ESCI supports baud rates between 110 and 19200 baud. The
default setting is 9600 baud.
BAUD RATE
19200
9600
4800
1200
600
300
110
IEC/IEEE-bus command: SYST:COMM:SER:BAUD 9600
Bits – Number of data bits per word
For the transmission of text without special characters, 7 bits are adequate.
For binary data as well as for text with special characters, 8 bits must be
selected (default setting).
BITS
7
8
IEC/IEEE-bus command: SYST:COMM:SER:BITS 7