Tracking Generator - Option R&S FSP-B9 R&S ESCI
1166.6004.12 4.276 E-1
CAL
TRANS
The CAL TRANS softkey triggers the calibration of the transmission
measurement.
It starts a sweep that records a reference trace. This trace is then used to
calculate the difference for the normalized values.
Fig. 4-26 Calibration curve for transmission measurement
During the calibration the following message is displayed:
After the calibration the following message is displayed:
This message will be cleared automatically after approx. 3 seconds.
IEC/IEEE-bus command: CORR:METH TRAN