R&S ESCI External Generator Control - Configuration
1166.6004.12 4.290 E-1
Transmission Measurement
The transmission characteristic of a two-port network is measured . The external generator serves as a
signal source. It is connected to the input connector of the DUT. The input of the analyser is fed from the
output of the DUT.
0
123
45
6
789
.
-
ESC
1142.8007.03
FCTN
ENTER
CANCE L
GEN OU TPUT 50
AF OUTPUT
MEAS TRI G
FREQ
MKR
AMPT
SPAN
MKR
MKR
BW SWEEP
TRACE
LINES
DISP
FILE
GHz
MHz
kHz
Hz
-dBm
dBm
dB
dB..
CAL
SETUP
PRESET
HCOPY
TEST RECEIVER 9kHz . . . 3GHz
..
ESPI
BACK
s
V
m
s
mV
µ
sµ
V
n
s
nV
PREV
RF INPUT
50
EXT MIXER
MA
X
+30 dBm / 50V DC
NEXT
KEYBOARDPROBEPOW ER
MAX 0V DC
MADE IN G ERMANY
.
RF
INPUT
GEN
OUTPUT
DUT
Fig. 4-34 Test setup for transmission measurement
A calibration can be carried out to compensate for the effects of the test setup (e.g. frequency response
of connecting cables).
Calibration of Transmission Measurement
NETWORK menu:
REF VALUE
POSITION
REF VALUE
SOURCE
CAL
NORMALIZE
CAL
TRANS
CAL REFL
OPEN
RECALL
CAL REFL
SHORT
The SOURCE CAL softkey opens a submenu comprising
the calibration functions for the transmission and reflection
measurement.
The calibration of the reflection measurement (CAL
REFL...) and its functioning are described in separate
sections.
To carry out a calibration for transmission measurements
the whole test setup is through-connected (THRU).