LAN Interface R&S ESCI
1166.6004.12 4.334 E-1
If the screen is dark or if a dark rectangle
appears in the upper left corner, the R&S
ESCI firmware must be restarted in order to
recognize the modified screen resolution.
In this case:
! Press <ALT><F4>.
The R&S FSP firmware will be shut down.
This may last for several seconds.
! Double-click the entry R&S Analyzer
Interface.
The firmware will restart and will
automatically open the Soft Frontpanel, i.e.
the user interface in which all front panel
buttons and the spinwheel are shown.
! If the R&S FSP application is directly
displayed after setup, shutdown and
restart can be omitted.
! All hardkeys, softkeys and hotkeys can be
operated with the mouse.
! The spinwheel is implemented by KNOB
LEFT, KNOB RIGHT and KNOB PRESS.
! The XP start menu is available if the
remote desktop window is enlarged to full
size.
Interruption and Re-setup of Remote Desktop Connection with the R&S ESCI
The connection to the R&S ESCI can be interrupted any time if the remote desktop window is connected
to the controller.
To reestablish the connection, the setup with the R&S ESCI only has to be performed again as
described in the section "Setting up the Connection with the R&S ESCI". The R&S ESCI will then
maintain its state in case of an interruption with a subsequent re-setup.
When the connection is set up with the controller, the login entry will be displayed on the R&S ESCI
screen. If the login procedure is carried out successfully, the message that a different user has assumed
control of the instrument and that the connection was therefore cleared will be displayed on the screen
of the controller.