R&S ESCI Example - Level and Frequency Measurement
1166.6004.12 2.15 E-2
10. From overview to standard-conformal
measurement
Data reduction and automatic routines for final
measurement
The R&S ESCI offers several data reduction
methods for interactive or automatic final
measurements.
The methods are described in Chapter 4, Section
"Data Reduction and Subrange Maxima".
Limit lines
A final measurement to standard can be performed
when active limit lines are displayed in the diagram
during the described analysis and measurements
are repeated at all critical frequencies using the
standard measurement time and a corresponding
detector.
The use and setting of limit lines is described in
Chapter 4, Section "Limit Lines - LIMIT Key".
Transducer
Care should be taken that the correction values or
transducer factors available in tabular form are
considered in the measurement result when
accessories with frequency-dependent
characteristics are used. Several correction tables
can be combined to form a transducer set.
The use and entry of transducer tables is described
in Chapter 4, Section "Using Transducers".
11. Storing test results, tables or diagrams on
floppy
In the example, test results are stored on a floppy.
Measured data are output to a file type *.wmf for use
in other applications.
DEVICE1 and output to a file type *.wmf is
automatically selected with PRESET.
The display elements are now selected.
HCOPY
! Press the HCOPY key.
The HCOPY SETTINGS menu is opened where the
output of measurement and device data can be
configured and started.
! Insert a floppy into the floppy disk drive.