The Screen R&S ESCI
1166.6004.12 3.6 E-1
Trace info:
Every active measurement curve (trace BLANK) is allocated trace
information of two or three lines at the left of the diagram. The trace
information has the same colour as the measurement curve.
The information on the currently selected trace is displayed in inverse
video (see TRACE - SELECT TRACE softkey).
<n> <detector> <*>
<mode>
<trace math>
Example:
1 PK *
CLRWR
1-2
n = trace number (1 ... 3)
detector = selected detector
AP: AUTOPEAK detector
PK: MAX PEAK detector
MI: MIN PEAK detector
SA: SAMPLE detector
AV: AVERAGE detector
RM: RMS detector
QP: QUASIPEAK detector
* = indicates that the selected detector does not correspond
to that of the automatic coupling.
mode = indication of sweep mode
CLRWR: CLEAR/WRITE
MAXH: MAX HOLD
MINH: MIN HOLD
AVG: AVERAGE
VIEW: VIEW
Trace math = trace math active
1 - 2 trace 1 - trace 2
1 - 3: trace 1 - trace 3