Common Analysis and Display Functions
R&S
®
ESW
217User Manual 1177.6298.02 ─ 08
Trace
Selects the trace the spectrogram is based on.
You can select any trace that is currently active, including the trace that shows the
results of trace mathematics.
Remote command:
CALCulate<n>:SPECtrogram:TRACe on page 532
Time Stamp
Activates and deactivates the timestamp. The timestamp shows the system time while
the measurement is running. In single sweep mode or if the measurement is stopped,
the timestamp shows the time and date of the end of the measurement.
When active, the timestamp replaces the display of the frame number.
Remote command:
CALCulate<n>:SPECtrogram:TSTamp[:STATe] on page 533
CALCulate<n>:SPECtrogram:TSTamp:DATA? on page 533
Color Mapping
Opens the "Color Mapping" dialog.
For details see "Color Maps" on page 208.
Clear Spectrogram
Resets the spectrogram result display and clears the history buffer.
This function is only available if a spectrogram is selected.
Remote command:
CALCulate<n>:SPECtrogram:CLEar[:IMMediate] on page 531
11.3.6.2 Color Map Settings
Access: "Overview" > "Analysis" > "Traces" > "Spectrogram" > "Color Mapping"
In addition to the available color settings, the dialog box displays the current color map
and provides a preview of the display with the current settings.