Network and Remote Operation
R&S
®
ESW
396User Manual 1177.6298.02 ─ 08
Language Comment
8566B
8568A Command sets A and B are available. Command sets A and B differ in the rules
regarding the command structure.
8568A_DC Uses DC input coupling by default if supported by the instrument
8568B Command sets A and B are available. Command sets A and B differ in the rules
regarding the command structure.
8568B_DC Uses DC input coupling by default if supported by the instrument
8591E Compatible to 8594E
8594E Command sets A and B are available. Command sets A and B differ in the rules
regarding the command structure.
PSA89600
PSA
Notes:
●
If you select a language other than "SCPI", the GPIB address is set to 18 if it was
20 before.
●
The Start/stop frequency, reference level and number of sweep points are adapted
to the selected instrument model.
●
When you switch between remote control languages, the following settings or
changes are made:
SCPI:
The instrument performs a PRESET.
8566A/B, 8568A/B, 8594E; FSEA, FSEB, FSEM; FSEK:
– The instrument performs a PRESET.
– The following instrument settings are changed:
Table 14-21: Instrument settings for emulation of 8566A/B, 8568A/B, 8594E; FSEA, FSEB, FSEM;
FSEK instruments
Model # of Trace
Points
Start Freq. Stop Freq. Ref Level Input Coupling
8566A/B 1001 2 GHz 22 GHz 0 dBm AC
8568A/B 1001 0 Hz 1.5 GHz 0 dBm AC
8560E 601 0 Hz 2.9 GHz 0 dBm AC
8561E 601 0 Hz 6.5 GHz 0 dBm AC
8562E 601 0 Hz 13.2 GHz 0 dBm AC
8563E 601 0 Hz 26.5 GHz 0 dBm AC
8564E 601 0 Hz 40 GHz 0 dBm AC
8565E 601 0 Hz 50 GHz 0 dBm AC
8594E 401 0 Hz 3 GHz 0 dBm AC
GPIB Languages