Remote Commands in the Receiver Application
R&S
®
ESW
453User Manual 1177.6298.02 ─ 08
15.5.5 Scan Configuration
Commands to configure scans described elsewhere.
See Chapter 15.5.6, "Scan Table Configuration", on page 454.
[SENSe:]FREQuency:MODE...........................................................................................453
[SENSe:]SCAN<sr>:TDOMain.........................................................................................453
[SENSe:]FREQuency:TDOPtim....................................................................................... 454
[SENSe:]FREQuency:MODE <Mode>
This command selects the scan mode.
Parameters:
<Mode> FIXed | CW
Selects fixed frequency scans.
Define the frequency with [SENSe:]FREQuency:CENTer.
SCAN | SWEep
Selects stepped scans in the frequency domain.
Define the frequency range with:
[SENSe:]FREQuency:STARt
[SENSe:]FREQuency:STOP
[SENSe:]SCAN<sr>:STARt
[SENSe:]SCAN<sr>:STOP
TDOMain
Selects time domain scans in the frequency domain.
Define the frequency range with the commands for stepped
scans.
*RST:
TDOMain
Example: //Select stepped scan
FREQ:MODE SCAN
Manual operation: See "Selecting the scan type" on page 113
[SENSe:]SCAN<sr>:TDOMain <Time>
This command defines the measurement time for fixed frequency scans.
Suffix:
<sr>
.
1..n
irrelevant
Parameters:
<Time> <numeric value>
Measurement time in seconds.
The range indicated below is the maximum range. The actual
range depends on the measurement time defined with [SENSe:
]SWEep:TIME on page 452.
Range: 10 ms to 10000 s
Default unit: s