Remote Commands in the Receiver Application
R&S
®
ESW
604User Manual 1177.6298.02 ─ 08
You have to use SYSTem:COMMunicate:PRINter:ENUMerate:FIRSt on page 603
for this command to work properly.
Manual operation: See "Printer Name" on page 285
SYSTem:COMMunicate:PRINter:SELect<1|2> <arg0>
This command selects the printer that processes jobs sent by the R&S ESW.
Use HCOPy:DESTination<1|2> to select another output destination.
Suffix:
<1|2>
.
1|2
Printing device.
Parameters:
<arg0> String containing the printer name.
Use
•SYSTem:COMMunicate:PRINter:ENUMerate:FIRSt
on page 603and
•SYSTem:COMMunicate:PRINter:ENUMerate[:NEXT]
on page 603
to query all available printers.
*RST: NONE
Manual operation: See "Printer Name" on page 285
15.8.5 Notes Display
Commands to configure the "Notes" display described elsewhere:
●
LAYout:ADD[:WINDow]?
DISPlay[:WINDow<n>][:SUBWindow<w>]:NOTes:APPend:TEXT........................................604
DISPlay[:WINDow<n>][:SUBWindow<w>]:NOTes:CLEar....................................................605
DISPlay[:WINDow<n>][:SUBWindow<w>]:NOTes:TEXT.....................................................605
DISPlay[:WINDow<n>][:SUBWindow<w>]:NOTes:APPend:TEXT <Text>
This command is used to append content to the notes display.
Suffix:
<n>
.
Window
<w> subwindow
Setting parameters:
<Text>
Example:
DISPlay:NOTes:APPend:TEXT 'Measurement
configuration'
'Measurement configuration' is added to the existing content of
the notes display.