Troubleshooting
R&S
®
ESW
725User Manual 1177.6298.02 ─ 08
When you select the error message bar or status message bar, the R&S ESW shows a
list of all current errors or status messages. In the error message bar, you can select
one of the error messages to open a dialog box that can help you to remedy the error.
Refer to the table below for information about which error message opens which dialog
box.
Table 17-2: List of keywords
AC COUP This label is displayed when a measurement is configured to be AC coupled, but
information in the DC range is detected. This situation can yield inaccurate measure-
ment results.
You can avoid this error message by changing the input coupling type.
Selecting the "AC Coup" error message opens a dialog box to configure the RF
input.
IF OVLD Overload of the IF signal path in the A/D converter or in the digital IF.
Increase the reference level.
Selecting the "IF OVLD" error message opens a dialog box to configure the ampli-
tude.
INPUT OVLD The signal level at the RF input connector exceeds the maximum.
The RF input is disconnected from the input mixer to protect the device. To re-enable
measurement, decrease the level at the RF input connector and reconnect the RF
input to the mixer input.
Selecting the "Input OVLD" error message opens a dialog box to configure the ampli-
tude.
LOUNL Error in the instrument's frequency processing hardware was detected.
NO REF Instrument was set to an external reference but no signal was detected on the refer-
ence input.
Selecting the "No Ref" error message opens a dialog box to configure the external
reference.
OVEN The optional OCXO reference frequency has not yet reached its operating tempera-
ture. The message usually disappears a few minutes after power has been switched
on.
RF OVLD Overload of the input mixer or of the analog IF path.
Increase the RF attenuation (for RF input).
Reduce the input level (for digital input)
Selecting the "RF OVLD" error message opens a dialog box to configure the ampli-
tude.
UNCAL One of the following conditions applies:
Correction data has been switched off.
No correction values are available, for example after a firmware update.
Record the correction data by performing a self alignment
Selecting the "Uncal" error message opens the "Self Alignment dialog box."
WRONG_FW The firmware version is out-of-date and does not support the currently installed hard-
ware. Until the firmware version is updated, this error message is displayed and self-
alignment fails.
Selecting the "Wrong_FW" error message opens a dialog box to manage firmware
and options.