Remote Commands in the Receiver Application
R&S
®
ESW
672User Manual 1177.6298.02 ─ 08
SYSTem:PRESet:FILTer <arg0>
This command selects the resolution filter type that is selected after a preset in the
Spectrum application.
Parameters:
<arg0> NORMal
Selects 3 dB filter.
NOISe
Selects 3 dB filter.
(NORMal and NOISe have the same effect.)
PULSe
Selects 6 dB filter.
*RST: NORMal
Example: //Select the 6 dB filters as the default filter type
SYST:PRES:FILT PULS
Manual operation: See "Default Filter Type for Spectrum Mode" on page 339
SYSTem:SECurity[:STATe] <State>
Activates or queries secure user mode.
Note: Before you activate secure user mode, store any instrument settings that are
required beyond the current session, such as predefined instrument settings, trans-
ducer files, or self-alignment data.
Note: Initially after installation of the R&S ESW-K33 option, secure user mode must be
enabled manually once before remote control is possible. This is necessary to prompt
for a change of passwords.
For details on the secure user mode see Chapter 4.7, "Protecting Data Using the
Secure User Mode", on page 40.
Parameters:
<State> ON | OFF | 0 | 1
ON | 1
The R&S ESW automatically reboots and starts in secure user
mode. In secure user mode, no data is written to the instru-
ment's internal solid-state drive. Data that the R&S ESW nor-
mally stores on the solid-state drive is redirected to SDRAM.
OFF | 0
The R&S ESW is set to normal instrument mode. Data is stored
to the internal solid-state drive.
Note: this parameter is for query only. Secure user mode cannot
be deactivated via remote operation.
*RST:
0
Manual operation: See "SecureUser Mode" on page 340
General Instrument Setup