Spare Part Replacement R&S FSH-Z2
1145.5767.02 3.6 E-1
Replacing the power cable assembly
(See exploded view for references).
½ Open the instrument as described earlier.
½ Remove the defective power cable assembly.
- Loosen screw (S4)
- Remove black shrink sleeve and desolder the two wires from the two pins.
- Remove the ferrite bead (item 06.1) and the black wire tule (item 06.2)
½
Fig. 3-6 Take the new power cable assy and slide the black wire tule (item 06.2) to 14.5 cm
½
Fig. 3-7 Put the wires through the ferrite bead (item 06.1).
½
Fig. 3-8 Solder the wires to the pins.