EasyManua.ls Logo

Rohde & Schwarz FSH13 - The Burst Power Result Display

Rohde & Schwarz FSH13
380 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
R&S FSH Digital Modulation Analyzer
Measurements on GSM Signals
Operating Manual 1173.6275.12 - 30 257
9.3.2 The Burst Power Result Display
If you equip the R&S FSH with option R&S FSH-K10E, the Burst Power result display
becomes available.
The Burst Power result display shows the power of the signal over time. You can
choose to display the power over a single and complete GSM frame (4.615 ms) or over
a single timeslot (about 567 µs)
This result display is designed to check if the power and timing of the bursted signals
meet the requirements of the standard.
In order to capture a complete frame, you can apply a trigger from a GPS device or
another external trigger.
Screen layout of the burst power result display
1 Current type of result display
2 Currently selected standard
3 Global results
4 Diagram area
5 Modulation type for each slot
6 GSM softkey menu

Table of Contents