R&S FSH Network Analyzer Mode
Performing Vector Measurements (R&S FSH-K42)
Operating Manual 1173.6275.12 - 30 194
Calibrate the measurement for vector transmission measurements.
Vector measurements provide the full range of calibration methods for more accurate
results. Most full calibration methods require more than one calibration standard.
Press the "Calibration" softkey.
Select the appropriate menu item (calibration on two ports).
Perform calibration. For more information see "Performing Calibration".
Reconnect the DUT.
The R&S FSH shows the results of the vector transmission measurement on the
filter.
The R&S FSH shows the results of
the vector transmission
measurement and therefore the
filter characteristics.
You can change the measurement
configuration (e.g. the sweep time
or detector) or format (e.g. view
phase characteristics) without
affecting the accuracy of the
measurement.
For more information see "Selecting the Measurement Format" on page 197.
Note that for calibration to remain valid, frequency parameters, bandwidth and
attenuation have to remain the same.
Depending on the calibration method, you can also select other result displays
(e.g. reflection measurements) without having to recalibrate the R&S FSH.