Loading...
R&S FSP Module Replacement
1164.4556.82 3.43 E-1
Replacing Diplexer (30 GHz) A230; R&S FSP13/30
(See chapter 5, spare parts list, item 150, and drawing 1164,4391.01.)
In R&S FSP13/30, the module is behind the attenuator on the right-hand side.
Removing Diplexer
Disconnect cable X71 from motherboard (515).
Disconnect cables W2 and W30 from diplexer (150).
Disconnect diplexer (150) from attenuator (30).
Remove diplexer (150).
Installing Diplexer
Screw on diplexer (150) to attenuator (30).
Fasten cables W2 and W30 to diplexer (150).
Connect cable X71 to motherboard (510).
Completing the Instrument and Frequency Response Correction
Execute frequency response correction software R&S FSP-FRQ.EXE.
The software is located on the Gloris server.
- Test setup according to menu item "Schematic" (connect external reference).
- Check IEC/IEEE addresses and configuration of instruments.
- Press the Autoselect button and perform the measurement with Run.
- The frequency correction of the YIG filter is performed under item YIG.
- After calibration, the frequency response is corrected over the entire frequency range.
Create a backup of the EEPROM data (see chapter 2, section "Adjustment of Module Data").