Preface
R&S
®
FSVA/FSV
11Quick Start Guide 1321.3066.02 ─ 06
Mixer settings for each band are saved and maintained even after preset-
ting the instrument; they can be preset explicitly using the "Preset Band"
function
If a second range (harmonic) is available, it can be activated and defined
freely, also for standard bands
When using two ranges for a band, the handover frequency can be defined
freely within the overlapping frequency range, also for standard bands
Additional traces are available for use when using the signal identification
functions Signal ID and Auto ID
All available conversion loss tables are offered for selection; validity tests
are performed after selection
1.5 Notes for Users of R&S FSV 1307.9002Kxx Mod-
els
Users of R&S FSV 1307.9002Kxx models should consider the following differen-
ces to the description of the newer R&S FSVA/FSV 1321.3008Kxx models:
Functions that are based on the Windows 10 operating system (e.g. printing
or setting up networks) may have a slightly different appearance or require dif-
ferent settings on the Windows XP based models. For such functions, refer to
the Windows documentation or the documentation originally provided with the
R&S FSV instrument.
The R&S FSV 1307.9002K03 model is restricted to a maximum frequency of
3 GHz, whereas the R&S FSVA/FSV1321.3008K04 model has a maximum
frequency of 4 GHz.
The bandwidth extension option R&S FSV-B160 (1311.2015.xx) is not availa-
ble for the R&S FSV 1307.9002Kxx models. The maximum usable I/Q analy-
sis bandwidth for these models is 28 MHz, or with option R&S FSV-B70,
40 MHz.
Notes for Users of R&S FSV 1307.9002Kxx Models