Remote Control
R&S
®
FSVA/FSV
788Operating Manual 1176.7510.02 ─ 09
This command is only available with option B21 (External Mixer) installed.
Parameters:
<SerialNo> Serial number with a maximum of 16 characters
Example:
CORR:CVL:SEL 'LOSS_TAB_4'
Selects the conversion loss table.
CORR:CVL:MIX '123.4567'
Mode: all
Manual operation: See "Mixer S/N" on page 530
[SENSe:]CORRection:CVL:HARMonic
<HarmOrder>
This command defines the harmonic order used in the conversion loss table. Before
this command can be performed, the conversion loss table must be selected (see
[SENSe:]CORRection:CVL:SELect on page 787).
This command is only available with option B21 (External Mixer) installed.
Parameters:
<HarmOrder> <numeric value>
Range: 2 to 65
Example:
CORR:CVL:SEL 'LOSS_TAB_4'
Selects the conversion loss table.
CORR:CVL:HARM 3
Mode: all
Manual operation: See "Harmonic Order" on page 530
[SENSe:]CORRection:CVL:PORTs <PortNo>
This command defines the mixer type in the conversion loss table. Before this com-
mand can be performed, the conversion loss table must be selected (see [SENSe:
]CORRection:CVL:SELect on page 787).
This command is only available with option B21 (External Mixer) installed.
Parameters:
<PortType> 2 | 3
*RST: 2
Example:
CORR:CVL:SEL 'LOSS_TAB_4'
Selects the conversion loss table.
CORR:CVL:PORT 3
Mode: all
Manual operation: See "Mixer Type" on page 530
Remote Control – Commands