6.3.1 Using Phase Noise Limit Lines
Phase Noise Limit Line................................................................................................. 59
Selected Traces............................................................................................................ 59
Noise Floor....................................................................................................................59
Range x - Range y........................................................................................................ 59
Copy to User Limit Line.................................................................................................60
Phase Noise Limit Line
Selects the shape of the phase noise limit line.
For more information see Chapter 4.9, "Using Limit Lines", on page 31.
"None"
No limit line.
"Noise floor
and x Ranges"
Limit line defined by the noise floor and x corner frequencies and
slopes. The application supports up to 5 ranges.
Remote command:
CALCulate:PNLimit:TYPE on page 107
Selected Traces
Selects the trace(s) to assign a phase noise limit line to.
For more information see Chapter 4.9, "Using Limit Lines", on page 31.
Remote command:
CALCulate:PNLimit:TRACe on page 107
Noise Floor
Defines the noise floor level in dBm/Hz of the DUT.
For more information see Chapter 4.9, "Using Limit Lines", on page 31.
Remote command:
CALCulate:PNLimit:NOISe on page 107
Range x - Range y
Defines the corner frequencies and slope for a particular segment of phase noise limit
lines.
The slope defines the slope of the limit line segment to the left of the corner frequency.
Using Limit Lines