147User Manual 1173.9240.02 ─ 17
Entries from 0 dB to 80 dB are allowed; the resolution is 0.1 dB. The default setting for
the peak excursion is 6 dB.
Remote command:
CALCulate<n>:MARKer:PEXCursion on page 308
Displaying Marker Numbers
By default, the marker numbers are indicated in the diagram so you can find the peaks
from the list. However, for large numbers of peaks the marker numbers may decrease
readability; in this case, deactivate the marker number display.
Remote command:
CALCulate<n>:MARKer<m>:FUNCtion:FPEaks:ANNotation:LABel[:STATe]
on page 314
Exporting the Peak List
The peak list can be exported to an ASCII file (.DAT) for analysis in an external appli-
cation.
Remote command:
MMEMory:STORe:PEAK on page 317
FORMat:DEXPort:DSEParator on page 292
6.7 Limit Line Settings and Functions
Up to 8 limit lines can be displayed simultaneously in the R&S FSW. Many more can
be stored on the instrument.
6.7.1 Limit Line Management...............................................................................................147
6.7.2 Limit Line Details.........................................................................................................150
6.7.1 Limit Line Management
Limit lines are managed in the "Line Config" dialog box which is displayed when you
press the LINES key and then "Lines Config" softkey.
Limit Line Settings and Functions