How to Perform Measurements in the Analog Demodulation Application
R&S
®
FSW-K7
162User Manual 1173.9240.02 ─ 17
Configure a trace to display the average over a series of sweeps (on the
"Trace" tab; if necessary, increase the "Sweep Count" in the "Data Acquisition"
settings).
Configure markers and delta markers to determine deviations and offsets within
the demodulated signal (on the "Marker" tab).
Use special marker functions to calculate phase noise or an n dB down band-
width (on the "Marker Config" tab).
Configure a limit check to detect excessive deviations (on the "Lines" tab).
10. Start a new sweep with the defined settings.
In multistandard mode you may want to stop the continuous measurement mode
by the Sequencer and perform a single data acquisition:
a)
Select the Sequencer icon (
) from the toolbar.
b) Set the Sequencer state to "OFF".
c) Press the RUN SINGLE key.
11. Optionally, export the trace data of the demodulated signal to a file.
a) In the "Traces" tab of the "Analysis" dialog box, switch to the "Trace Export"
tab.
b) Select "Export Trace to ASCII File".
c) Define a file name and storage location and select "OK".

Table of Contents

Other manuals for Rohde & Schwarz FSW Series

Related product manuals