Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
193User Manual 1173.9240.02 ─ 17
This command is only available with option B21 (External Mixer) installed.
Example:
CORR:CVL:SEL 'LOSS_TAB_4'
Selects the conversion loss table.
CORR:CVL:CLE
Usage: Event
Manual operation: See "Delete Table" on page 65
[SENSe:]CORRection:CVL:COMMent <Text>
This command defines a comment for the conversion loss table. Before this command
can be performed, the conversion loss table must be selected (see [SENSe:
]CORRection:CVL:SELect on page 195).
This command is only available with option B21 (External Mixer) installed.
Parameters:
<Text>
Example:
CORR:CVL:SEL 'LOSS_TAB_4'
Selects the conversion loss table.
CORR:CVL:COMM 'Conversion loss table for
FS_Z60'
Manual operation: See "Comment" on page 67
[SENSe:]CORRection:CVL:DATA <Freq>,<Level>
This command defines the reference values of the selected conversion loss tables.
The values are entered as a set of frequency/level pairs. A maximum of 50 frequency/
level pairs may be entered. Before this command can be performed, the conversion
loss table must be selected (see [SENSe:]CORRection:CVL:SELect
on page 195).
This command is only available with option B21 (External Mixer) installed.
Parameters:
<Freq> numeric value
The frequencies have to be sent in ascending order.
<Level>
Example:
CORR:CVL:SEL 'LOSS_TAB_4'
Selects the conversion loss table.
CORR:CVL:DATA 1MHZ,-30DB,2MHZ,-40DB
Manual operation: See "Position/Value" on page 68