Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
204User Manual 1173.9240.02 ─ 17
<SampleRate> Maximum or currently used sample rate of the connected device
in Hz (depends on the used connection protocol version; indica-
ted by <SampleRateType> parameter)
<MaxTransferRate> Maximum data transfer rate of the connected device in Hz
<ConnProtState> State of the connection protocol which is used to identify the
connected device.
Not Started
Has to be Started
Started
Passed
Failed
Done
<PRBSTestState> State of the PRBS test.
Not Started
Has to be Started
Started
Passed
Failed
Done
<SampleRateType> 0
Maximum sample rate is displayed
1
Current sample rate is displayed
<FullScaleLevel> The level (in dBm) that should correspond to an I/Q sample with
the magnitude "1" (if transferred from connected device);
If not available, 9.97e37 is returned
Example:
INP:DIQ:CDEV?
Result:
1,SMU200A,103634,Out
A,70000000,100000000,Passed,Not Started,0,0
Manual operation: See "Connected Instrument" on page 70
INPut:DIQ:RANGe[:UPPer]:AUTO <State>
If enabled, the digital input full scale level is automatically set to the value provided by
the connected device (if available).
This command is only available if the optional Digital Baseband interface (option
R&S FSW-B17) is installed.
Parameters:
<State> ON | OFF
*RST: OFF
Configuring the Measurement